Spectrophotometer & Beer’s Law Lab

AP Chemistry
File ▾

AP CHEMISTRY · UNIT 3

How much dye is in the solution? Ask the light.

A spectrophotometer shines one colour of light through a solution and measures how much is absorbed. The more dye in the light’s path, the more light it absorbs: that is Beer’s law, A = εbc.

  1. Zero the instrument with a blank (a cuvette of water).
  2. Put a sample in, choose a wavelength, and read the absorbance.
  3. Record the reading, then change one thing and repeat.

Choose where to start:

About this model

The law. Absorbance A = εbc (the Beer–Lambert law) and A = −log₁₀(T), where T is the fraction of light that passes through. b is the path length, 1.00 cm in this cuvette, and c is the concentration. Absorbance has no unit.

The dye. FD&C Blue No. 1 (brilliant blue), with ε = 130 000 L mol⁻¹ cm⁻¹ at λmax ≈ 630 nm (the values used in Vernier’s AP Chemistry food-dye lab). The standards are 2.0 to 10.0 µmol/L, so absorbances run from about 0.26 to 1.30.

What is simplified. The shape of the absorbance spectrum is a smooth bell curve chosen for teaching, not a measured spectrum: the real dye’s band is narrower with a shoulder near 590 nm. The model is an ideal dilute solution of one dye: perfectly linear, with no stray light, no scattering and no drift. Real calibration lines bend at high absorbance.

The readings. Each reading carries a small random error of up to ±0.003 absorbance, the same at every wavelength, so repeat readings differ in the last digit. That size is illustrative; check your own instrument’s manual. Bench questions in the test show exact model values.

The blank. Like a classroom diode-array spectrophotometer, one zero with the blank stores a baseline at every wavelength. On an older single-wavelength instrument you zero again each time you change the wavelength.

The colours. The beam colour, the colour of each solution and the “light that gets through” band are computed from the model’s spectrum with the standard (CIE 1931) colour-matching functions, then shown as closely as a screen allows. Pure spectral colours cannot be shown exactly on a screen. The instrument picture is a schematic and is not to scale.

What am I looking at?

How the lab works. Pick Explore or one lab at the top. Each lab has five tabs: Pre-lab, Bench, Data & graph, Claim · Evidence · Reasoning, and Test. On the Bench, the coach shows your next step; steps tick themselves as you work. Press What do I do now? when you’re stuck.

The instrument. The picture is a cutaway of a spectrophotometer. White light from the light source is spread into a spectrum by a grating; a slit lets one narrow band through (together, the monochromator). That beam crosses 1.00 cm of solution in the cuvette and the detector measures what is left. The green display shows the wavelength, the absorbance and the percent transmittance.

The two colour bands. The top band is the whole visible spectrum, with a pointer at the dial’s wavelength. The lower band is the same spectrum after it has passed through the cuvette: dark places are light the solution absorbed. The colour you see is made of the light that is left.

Controls. Drag the wavelength dial (or tap the spectrum band; arrow keys and the ± buttons move 1 nm). Tap a cuvette to put it in the holder; tap it again to take it out. Zero with blank sets the blank to absorbance 0. Record reading adds the display’s reading to your data table.

The graph. Next to the bench: absorbance against wavelength (a spectrum), or absorbance against concentration (a calibration). Filled circles are readings you recorded; the hollow diamond is the reading on the display now.